Wheelchair scale is a product specially tailored for the medical industry. It adopts a modern and simple appearance design to enhance the grade of hospital units. The bright LED display
1. Model: CI-2001ASH weighing indicator
2. Grade: III grade division number n = 3000
3. Input signal: 1.0 2.0 mV / V
4.A / D conversion speed: 40 times / second
5. Bridge voltage: 5V DC can drive 4 350W or 8 700W sensors
6. A / D resolution code: 300,000
7. Voltage: 220V 50HZ
8. Built-in: 6V / 10Ah maintenance-free battery, can work continuously for 24 hours after charging
9. Working temperature: -10 40
10. Working humidity: <85% RH
§1. Power switch Press the [] key to turn on the power. Press the [] key for 2 seconds to cut off the power.
§2. Zeroing Press and hold the [ 0 ] key for 2 seconds, the displayed value will be set to zero, and the [ 0 ] indicator will light up.
§3. Peeling
§3.1 The peeling of the real object is in the state where the indicator light is off, and the weighing is greater than zero and stable. Press the [ 0 ] key to set the display value to zero.
At the same time, the indicator lights up.
§3.2 Clearing tare weight The indicator light is on. Press the [ 0 ] key to clear the tare weight. At the same time, the indicator light is off.
§3.3 Digital Tare In the state where digital tare is allowed, press the [] key to enter the tare weight setting, and use the [] [] keys
Enter the tare value and press the [ 0 ] key to store the input value as the tare weight. At the same time, the indicator lights.
§4. The accumulation operation is in the manual accumulation state. If the weighing is greater than zero and stable, press to accumulate the displayed value and display
The cumulative number is displayed. [n 12].
Note: One weighing can only be accumulated once. The next accumulation must be allowed after the displayed value is less than 20d.
§5. The high-resolution display is in the weighing display state. Press the [] key to enter the high-resolution display (10 times increase).
The decimal point lights up. Press the [] key again to return to the normal weighing display.
§6. Display and clear of accumulated value
Press the [] key to display the total number of times [n 12]. After 2 seconds, the total weight is displayed [1236.5]
If the weight is more than 6 digits, the accumulated weight is displayed 4 digits higher [H 123], and the accumulated weight is displayed 4 digits after 2 seconds
Bit [L 506.5], (total weight is 123506.5) In the above display process, press the [ 0 ] key,
Clear the accumulated value.
Press the [] key to display the automatic accumulation status, [Aut 0].
Last bit = 0, means manual accumulation
Last bit = 1, it means automatic accumulation after stable loading
Last bit = 2, which means that the display value is automatically saved after the loading is stable.
After unloading to less than 20d, accumulate the final stable value
Use [] to set and press [] to return to normal weighing display.
Note: In the automatic accumulation state, the [AUT] indicator is on.
§7. The digital input [] key is used to move the setting digit, and the setting digit blinks. [] key is used to increase the value of the set digit.
§8. When the battery power is on and off, [Pbt 85] is displayed, indicating that the battery power is 85%.
§9. Charging reminder When the battery level is less than 15%, the display will flash to prompt prompt charging.
§10. Over-discharge protection When the battery level is zero, the meter will automatically shut down to prevent over-discharge of the battery.
§11. If the power-saving mode is set to have an automatic shutdown function, and there is no weighing operation for more than 3 minutes, [-] is displayed to reduce power
Pool consumption.
§12. Automatic shutdown If the automatic shutdown function is set, if there is no weighing operation for more than 15 minutes, it will automatically shut down.
§13. Install 10.0AH maintenance-free battery during working hours, sufficient battery, can work continuously for 24 hours.
Calibration and weighing setting operation
Only when the calibration allows the socket to be on the "EN" side, the version shows [UE-0.06]. The operation is valid.
In the shutdown state, press and hold the [ 0 ] key and do not let go. Press the [] key to turn on the power. The "-" at the end of the version indicates that calibration is allowed.
step
operating
display
annotation
1
press*
[CAL SP]
Weighing calibration, if you want to enter weighing calibration, press [], otherwise press []
2
press*
[-Set-]
Weighing parameter setting, if you want to enter the parameter setting, press [], otherwise press []
3
press*
[-Ad-]
A / D value display, if you want to enter A / D value display, press [], otherwise press []
4
press*
[15.0]
Return to weighing status
§1. Weighing settings
step
operating
display
annotation

press*
[CAL SP]
Weighing calibration

press*
[-Set-]
Weighing parameter settings, if you want to enter, press []

Press []

Enter weighing parameter setting


[d 0.1]
Division value
1
Press []
0.0001--50
Select one of the division values
2
press*
[300.0]
Maximum weighing
3
Press [] []

Enter new maximum weight
4
press*
[g X]
Zero tracking range
5
Press []

Enter new value (range = 0.3 d * X)
6
press*
[3.0]
Minimum weighing
7
Press [] []

Enter new minimum weighing value
8
press*
[F edcba]
Function settings



e = 0, no automatic shutdown function



e = 1, with automatic shutdown function, no weighing operation for more than 3 minutes,
[] Power saving mode, automatic shutdown after 15 minutes



d = 0, no unit switching function



d = 1, unit switching function is allowed



c = 1, allow digital tare



b = 0, no accumulation function



b = 1, allow accumulation function



a = 0, no high-resolution display function



a = 1, high-resolution display function is allowed
9
Press [] []

Setting of each function
10
press*
[FLt X]
Filtering parameter settings



X = 0, static weighing method, widely used in general weighing



X = 1, the weighing method for crane scales, used for weighing cranes and other hanging objects



X = 2, animal weighing method, applied to weighing objects in non-stationary state
Note: In animal weighing mode: When the weight is greater than the minimum weight and is stable, the display is locked; when the weight is less than the minimum weight, the display is unlocked.



11
Press []

Selecting filtering parameters
12
press*
[Adr XX]
XX means communication address
13
Press [] []

Set communication address



X = 0, continuous sending mode, 6 times / second



X = 1 ~ 98, response mode



X = 99, printer with serial interface
14
press*
[b 9600]
Communication baud rate
15
Press []

Select communication baud rate 1200, 2400, 4800, 9600
16
press*
[-Ad-]
A / D value display, if you want to enter A / D value display, press [], otherwise
17
press*

Return to weighing status
§2. Calibration
Only when the calibration allows the socket to be on the "EN" side, the version shows [UE-0.06] operation is valid
step
operating
display
annotation

press*
[CAL SP]
Weighing calibration, if you want to enter, press []
1
Enter weighing calibration
[CAL OO]
Enter zero calibration
2
press*
Press [] for 2 seconds
[-------]
Zero calibration
Switching the calibration unit, the [kg] light is off, and the [lb] light is on
3
After a few seconds
[300.0]
Zero calibration completed, maximum weighing
4
Press [] []
[150.0]
Load weight, enter the load weight value
5
press*
[-------]
Span calibration
6
After a few seconds
[150.0]
The calibration is completed, the loaded weight value is displayed, and the weighing state is entered.
Note: 1 (compensation calibration)
If the zero position is not changed, only the span calibration is performed. When [CAL OO] is displayed, press [ 0 ] to skip the zero calibration and enter the span calibration.
Note: 2 (zero calibration)
If only zero calibration is performed, then when the zero calibration is completed and the maximum weighing value [300.0] is displayed, use the [] and [] keys to enter zero.
[0.0], press [] to end the calibration and return to normal weighing display.
§3. After calibration is completed, shut down. The calibration allows the socket to be on the "DS" side, and then power on.
§4. A / D value display A / D value display is used to check whether the meter or sensor is working properly.
Generally, the no-load A / D value is around 40,000.
§5. Sensor connection (D-SUB-9 hole)
1-2 pins + E (positive sensor excitation) supply bridge voltage 5VDC current 70 mA
3-pin GND (sensor shield ground) sensor sensitivity 1.0-2.0 mV / V
4-5 pin-E (sensor negative excitation) A / D conversion speed 40 times / second
7 pin + S (positive sensor output) A / D resolution 100000
8-pin-S (sensor negative output) AC 220W / 50Hz
§6. Communication connection (D-SUB-9 pin)
2-pin RXD (RS-232 input)
3-pin TXD (RS-232 output)
5 pin GND
§7. Communication data format
Communication address X = 0; continuous sending mode, format is
STX X1 X2 X3 X4 X5 X6 SA CR
X1 X2 X3 X4 X5 X6 is display data SA = 0100ABCD is status information
A = 1; Weighing overflow B = 1; Tare C = 1; Weighing stability D = 1; Zero position
Communication address X 0; response mode
Read weight command: STX RDS n BCC CR
n = communication address (81H E2H) BCC is the parity bit
Returned data: STX n D X1 X2 X3 X4 X5 X6 SA BCC CR
Peeling and peeling commands: STX TAR n BCC CR
n = communication address (81H E2H) BCC is the parity bit
Returned data: STX n D X1 X2 X3 X4 X5 X6 SA BCC CR
BCC is the check code. The sum of data from STX to BCC is taken as the lower 8-bit data, but when the value is equal to 02 or 0D, it is automatically increased by 1 (hexadecimal number) to avoid the control character (STX = 02; CR = 0D) Repeat.
Chapter 4 Error Tips and Fault Diagnosis
1. Err E: indicates that the instrument memory is faulty.
2. Err 1: It means that the A / D value is lower than the normal range, maybe the meter or the sensor is faulty.
Disconnect the sensor from the meter, the meter error display disappears, indicating that the sensor is faulty.
Disconnect the sensor from the meter, the meter error display does not disappear, indicating that the meter is faulty.
3. Err 2: It means that the A / D value is higher than the normal range. The meter or sensor may be faulty.
Unplug the sensor, and the error display disappears, indicating that the sensor is faulty.
Unplug the sensor, and the error display of the instrument will not disappear, indicating that the instrument is faulty.
4 Err 3: It indicates that the zero position at startup is abnormal.
5. Err 4: It means that the A / D is too small during zero calibration.
6. Err 5: The span calibration has failed.
7. Err 6: It means that the weighing weight cannot be stable during calibration.
8. [OUEr]: Indicates that the weighing weight is greater than the maximum weighing.
Disconnect the sensor from the meter, the meter error display disappears, indicating that the sensor is faulty.
Disconnect the sensor from the meter, the meter error display does not disappear, indicating that the meter is faulty.
Enter A / D value display to further check whether the meter works normally
If the A / D works normally, the maximum weighing setting may be incorrect. Can be reset.
Or recalibrate the scale to resume normal work.
9. Weighing instability: The meter or sensor may be faulty.
Disconnect the sensor from the meter and weigh it steady, indicating that the sensor is faulty.
Disconnect the sensor from the meter, weighing the unstable weight, indicating that the meter is faulty.
10. Cannot turn on the machine: The battery in the meter may be empty, or the sensor may be short-circuited, or the meter may be faulty.
Disconnect the sensor from the meter and turn it on, indicating that the sensor is shorted.
External power supply, AC indicator is off. Indicates that the meter is faulty.
External power supply, AC indicator is on. But the machine cannot be turned on, indicating that the instrument is faulty.