What is Compartment Syndrome?
Pre-excitation alone is asymptomatic. Concurrent supraventricular tachycardia is similar to general supraventricular tachycardia. In patients with atrial flutter or atrial fibrillation, the ventricular rate is about 200 beats / min. In addition to palpitations and other discomforts, shock, heart failure and even sudden death may occur. When the ventricular rate is extremely fast, such as 300 beats / min, auscultatory heart sounds can only be half of the ventricular rate on the ECG, suggesting that half of ventricular agitation cannot produce effective mechanical contraction.
Preexcitation syndrome
- Pre-excitation is an abnormal phenomenon of atrioventricular conduction. Impulse is transmitted through the additional channel, which excites part or all of the ventricle early, causing some ventricular muscles to excite in advance. Those with pre-excitation are called pre-excitation syndrome or WPW (Wolf-Parkinson-White) syndrome, often with supraventricular paroxysmal tachycardia. Pre-excitation is a rare type of arrhythmia, and diagnosis is mainly based on electrocardiograms.
Preexcitation Syndrome
- Pre-excitation alone is asymptomatic. Concurrent supraventricular tachycardia is similar to general supraventricular tachycardia. In patients with atrial flutter or atrial fibrillation, the ventricular rate is about 200 beats / min. In addition to palpitations and other discomforts, shock, heart failure and even sudden death may occur. When the ventricular rate is extremely fast, such as 300 beats / min, auscultatory heart sounds can only be half of the ventricular rate on the ECG, suggesting that half of ventricular agitation cannot produce effective mechanical contraction.
Causes of Preexcitation Syndrome
- Congenital atrioventricular additional channels (referred to as bypass) exist outside the normal atrioventricular conduction system. Most patients have no organic heart disease. It is also found in certain congenital and acquired heart diseases, such as tricuspid valve downward movement and obstructive cardiomyopathy. Electrophysiological research proves that the bypass has a fast conduction speed, and the atrial impulses pass down the bypass quickly, reaching the ventricular end of the bypass in advance, and excite the adjacent myocardium, which causes the ventricles to excite in advance and change the normal excitation sequence of the ventricular muscle. The QRS complex on the electrocardiogram is deformed with a pre-shock ( wave) at the beginning. The rest of the atrial impulse can pass down the normal path, and merge with the ventricular excitement caused by the bypass to form a ventricular fusion wave. The shape of the ventricular fusion wave is determined by the refractory period of normal and bypass. The normal pathway should not have a long period, or most impulses are conducted along the bypass, the QRS deformity is obvious; the long bypass should not, the ventricular fusion wave is close to normal.
- There are two conduction pathways in the atrioventricular compartment of patients with preexcitation syndrome, which are prone to reentry and reentrant tachycardia. During tachycardia, most of them pass through the bypass and pass down the normal channel, so the QRS complex of tachycardia is normal. Occasionally, impulses pass through the bypass and pass back through the normal channel, causing QRS when tachycardia occurs. The wave group is pre-excited. Pre-excitation patients may also have atrial fibrillation or atrial flutter attacks, which are mostly caused by impulsive retrograde transmission and reaching the atrium during the vulnerable period of the atrium. In atrial flutter and atrial fibrillation, the impulse conducts in the tissue at the junction, prompting most or all of the impulses to pass to the ventricle. Atrial flutter or atrial fibrillation with extremely fast ventricular rate and QRS complex may sometimes develop into ventricular fibrillation. The one-way block of bypass (mostly the down-pass block) can make the ECG without preexcitation, but it has recurrent episodes of supraventricular tachycardia; electrophysiological studies can confirm that the bypass is involved in the return of tachycardia. The second degree conduction block of the bypass can cause intermittent appearance of pre-excitation on the ECG. The following types of bypasses are known, and multiple bypasses can be used for the same patient: atrioventricular bypass (Kent bundle). Mostly located on the left and right sides of the atrioventricular sulcus or the septum, connecting the atrial muscles and ventricular muscles; atrial node pathway (James pathway). It is the channel between the atrium and the lower part of the atrioventricular node or the atrioventricular bundle, which may be formed by the fibers of the posterior internode bundle; the nodule and the bundle chamber connection (Mahaim fiber). It is the pathway connecting the distal end of the atrioventricular node or the proximal end of the atrioventricular bundle or bundle branch to the ventricle. Atrioventricular by-pass is the most common of the three.
Clinical diagnosis of preexcitation syndrome
Pre-excitation syndrome without pre-excitation is asymptomatic
- Concurrent supraventricular tachycardia is similar to general supraventricular tachycardia
- Performance of different bypasses from anatomy, electrocardiogram, and Sis bundle
ECG vector diagram diagnosis of preexcitation syndrome
- In addition to the features of the electrocardiogram, the electrocardiogram vector diagram can be used as a diagnostic basis, which is characterized by the slow start of the QRS ring on each face running in a straight line, which lasts for 0.08 seconds, and then suddenly turns and continues to run at normal speed. The QRS ring run time can exceed 0.12 seconds. Heath beam electrocardiogram and body surface or epicardial mapping help identify various types of pre-excitation and localization of the bypass, and play an important role in determining whether the bypass is involved in the tachycardia reentry loop. The pre-excitation pattern on the ECG should be distinguished from bundle branch block, ventricular hypertrophy or myocardial infarction. Shortening of the PR interval and the presence of pre-excitation waves can be confirmed as pre-excitation. When the accelerated ventricular spontaneous rhythm and sinus rhythm are interfering with atrioventricular separation (especially when the ventricular rate is similar to the sinus heart rate), short-term PR intervals may be shortened, and the QRS complex may be broadly deformed. Sexual pre-excitation; but long records can often show that the PR interval is not fixed and the atrioventricular separation is not difficult to distinguish from pre-excitation. When pre-excitation is accompanied by supraventricular tachycardia, the QRS complex is often not widened, but after the onset of interruption, there are characteristic ECG changes other than occult pre-excitation. When pre-excitation is accompanied by atrial fibrillation or atrial flutter, the QRS complex is often Widening should be distinguished from ventricular tachycardia.
Pre-excitation syndrome test
- ECG performance
- The characteristics of the electrocardiogram caused by pre-excitation of each bypass are as follows.
- (1) The atrioventricular path The PR interval (essentially the P- interval) is shortened to less than 0.12 seconds, mostly 0.10 seconds; The QRS time limit is extended to more than 0.11 seconds; The beginning of the QRS complex is blunt, and the rest Formation of a frustration, the so-called pre-excitation secondary ST-T wave changes.
- Preexcitation syndrome
- The ECG changes mentioned above can be divided into A and B types. Both the pre-excitation wave and QRS complex of type A are upward in lead V1 (Figure 3), while the main wave of pre-excitation wave and QRS complex of lead B of type V1 are both downward; the former indicates the left ventricle or right ventricle. Posterior bottom myocardial preexcitation, which in turn suggests anterior ventricular myocardial premuscle. Although this classification method is limited by the pre-excitation caused by the variable QRS complexes of the bypass in different parts, it helps to distinguish the ventricular end of the bypass from left or right, anterior or posterior, and is still used today.
- Type A pre-excitation syndrome shows V1 pre-excitation and QRS complex main wave upward
- (2) The PR interval of the atrial node and Fangxi side road is less than 0.12 seconds, most of which are 0.10 seconds; the QRS complex is normal and there is no pre-shock. This ECG manifestation is also known as short PR, normal QRS syndrome, or L, G, L (Lown-Ganong-Levine) syndrome.
- (3) The PR interval between the nodal chamber and the beam chamber is normal, the QRS complex is widened, and there are pre-shocks.
- Pre-excitation tachycardia onset of pre-excitation syndrome, the pre-excitation manifestation mostly disappeared, and the electrocardiogram showed a QRS complex with supraventricular tachycardia (Figure 5). It is not uncommon for QRS to maintain pre-excitation characteristics during concurrent atrial flutter or atrial fibrillation (Figure 6). The ECG manifests a wide atrial flutter or atrial fibrillation with a QRS complex; the ventricular rate mostly exceeds 200 beats / min, or even 300 / min. Atrial flutter can present 1: 1 atrioventricular conduction and may identify atrial flutter waves. The ventricular rhythm is irregular during atrial fibrillation, and individual QRS complexes can be seen to be normal after a long interval (probably the bypass refractory period is prolonged, after the hidden conduction in the atrioventricular node disappears, all or most of the impulses pass through the atrioventricular node Cause) and may identify atrial fibrillation. When the ventricular rate is extremely fast, it can also be accompanied by frequency-dependent changes in intraventricular conduction.
- Pre-excitation syndrome with supraventricular tachycardia. Left: Sinus rhythm pre-excitation performance. Right: Supraventricular tachycardia onset. QRS complex morphology is normal.
- Pre-excitation comprehensive film with atrial fibrillation Left: Atrial fibrillation onset, ventricular rate 167-300 beats / min, average 210 beats / min, significantly irregular Right: sinus rhythm, with typical pre-excitation performance.
Pre-excitation Syndrome Treatment
- Preexcitation syndrome
- If the QRS wave is normal, the P-R interval is regular, and the heart rate is about 200 beats / min. It should be considered as repetitive tachycardia. The treatment is the same as that of general supraventricular tachycardia. , ATP, or digitalis, if the QRS complex is abnormal and the R-R interval is significantly irregular, it should be suspected and pre-excited with atrial fibrillation, you should choose cardiac rhythm, procainamide, or quinidine and experience An combined use, while isopadine, digitalis and ATP, because the latter three can shorten the bypass refractory period and accelerate bypass conduction, and even ventricular fibrillation occurs.
- For patients with frequent episodes of supraventricular tachycardia and obvious symptoms, electrophysiological examination should be performed to identify the side-channel site and then use electrical ablation, radiofrequency ablation, or surgical treatment.
Pre-excitation Syndrome Care
Theoretical analysis of pre-excitation syndrome
- Preexcitation syndrome
Pre-excitation syndrome prevention strategy
- Preexcitation syndrome
- If you do nt use ECG, it is very difficult. Because of the different types, the consequences will be very different. Some jumps are not dangerous for a few days, while others are about to die in a few minutes. You must fight against the clock every second, so when tachycardia occurs Need to do an electrocardiogram immediately. On the electrocardiogram, the doctor can determine what tachycardia is, which part of the heart is tachycardia, and how dangerous it is, so as to achieve effective symptomatic treatment and symptomatic treatment. [1]